INSTALLATION INFORMATION
• Plumbing (water) lines
To assure that the SUPERB VAC provides optimum Vacuum, incoming water pressure must be maintained between 20 and 100 psi
Always TURN ON THE WATER before TURNING ON THE POWER
If heavy combinations of particulates exist in the incoming water, an in-line filter should be installed. This will prevent the SUPERB VAC's water inlet filter from clogging too frequently
Incoming water temperature should be between 40° and 75°F
Water connection location is shown in KEY PARTS IDENTIFICATION
• Suction
The suction hose is connected at the suction intake
• Drain lines
For Superb Vacs with an A/W Separator, gases should be vented out according to code. The wastewater (with particulates) from the operatories can be discharged via an open drain or a closed vented drain.
• Electric
To help prevent fire, electric shock, injury, or death, the wiring and grounding must conform to the latest edition of the National Electrical Code, ANSI/NFPA70, and all applicable local regulations. Please contact a qualified electrician to check your wiring and breakers/ fuses to ensure that there is adequate electrical power to operate the Vacuum pump.
All SUPERB VAC units must be connected to a grounded metal, permanent wiring system, or an equipment grounding conductor must be run with the circuit conductors and connected to the equipment- Grounding/Protective Earth lead on the SUPERB VAC pump’s flexible metal conduit power supply. Failure to do so can result in fire, electric shock, injury, or death.
All SUPERB VAC are wired with a supplied hospital-grade NEMA 6-15P line cord and requires a hospital-grade 6-15R receptacle. The Superb Vac is wired with a supplied hospital-grade NEMA 6-15P line cord and requires a grounded hospital-grade 6-15R receptacle located near the unit Outlet must meet all local codes and ordinances.
All SUPERB VAC units must be connected to a grounded metal, permanent wiring system, or an equipment grounding conductor must be run with the circuit conductors and connected to the equipment grounding lead in the SUPERB VAC’s flexible metal conduit power supply. Failure to do so can result in fire, electric shock, injury, or death.